11
Measurement tab:
Performed measurements can be viewed in the “Measurements” tab.
If several measurements have already been carried out, select the desired ABI
measurement in the “Date-time” field.
The following parameters will be displayed on the screen from top to bottom (two
columns – one for each side of the body):
DiffSys Arm (difference between the systolic values of the left and the right upper
arm)
DiffDia Arm (difference between the diastolic values of the left and the right upper
arm)
Sys (systolic blood pressure of the upper arm), displayed in red > 140 mmHg
Dia (diastolic blood pressure of the upper arm), displayed in red> 90 mmHg
PP (pulse pressure = difference between systole and diastole), displayed in red
>54 mmHg
Pul (pulse value) in 1/min.
Arr (indicates whether a pulse frequence anomaly of more than 25% occurred
during a measurement (“Yes”, displayed in red, or “No”).
ABI (ankle-brachial index = quotient from the leg measurement’s systolic value
divided by the higher systolic value from both upper arm measurements), by def-
ault displayed in red < 0.9.
Sys (ankle systolic blood pressure)
Note: If no measurement result is displayed for the ankle’s systolic blood pressure,
this may indicate a circulatory disturbance or media sclerosis. Further investigati-
on is recommended.
In addition:
ΔSys(differencebetweenthesystolicvalueoftheleftupperarmandtheright
upper arm).
ΔDia(differencebetweenthediastolicvalueoftheleftupperarmandtheright
upper arm).
ΔSys(differencebetweenthesystolicvalueoftheleftandtherightankle)